DC/DC Medical Converter THM 6WI Series, 6 Watt Ultra wide 4:1 input voltage 6 W DC/DC converter in a compact DIP-24 plastic case I/O isolation 5000 VAC rated for 250 VAC working voltage Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP ES 60601-1 IEC 60601-1 UL 62368-1 IEC 62368-1 Risk management process according to ISO 14971 incl. risk management file The THM 6WI series is a range of medical 6 Watt DC/DC converters in DIP-24 plastic package and with ultra-wide 4:1 input voltage range. They provide a re- Acceptance criteria for electronic inforced isolation system for 5000 VAC isolation and a very low leakage current assemblies acc. to IPC-A-610 Level 3 of less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd edition Low leakage current <2 A for 2 x MOPP and come along with an ISO 14971 risk management file. Design and production conform to the quality management system ISO 13485. With a Operating temperature 40C to 90C high efficiency of up to 87% and highest grade components the converters can EMC compliance to IEC 60601-1-2 4th reliably operate in an ambient temperature range of 40C up to +90C. They edition and EN55032 class A constitute a reliable solution not only for medical equipment but also for de- manding ranges of application such as transportation, control & measurement Operating up to 5000m altitude or IGBT drivers. 5-year product warranty Models Order Code Input Voltage Output 1 Output 2 Efficiency Range Vnom Imax Vnom Imax typ. THM 6-0510WI 3.3 VDC 1 800 mA 82 % THM 6-0511WI 5 VDC 1 200 mA 86 % THM 6-0512WI 12 VDC 500 mA 86 % THM 6-0513WI 4.5 - 9 VDC 15 VDC 400 mA 88 % (5 VDC nom.) THM 6-0515WI 24 VDC 250 mA 87 % THM 6-0521WI +5 VDC 600 mA 5 VDC 600 mA 84 % THM 6-0522WI +12 VDC 250 mA 12 VDC 250 mA 87 % THM 6-0523WI +15 VDC 200 mA 15 VDC 200 mA 88 % 3.3 VDC 1 800 mA 83 % THM 6-2410WI THM 6-2411WI 5 VDC 1 200 mA 86 % THM 6-2412WI 12 VDC 500 mA 89 % THM 6-2413WI 9 - 36 VDC 15 VDC 400 mA 89 % (24 VDC nom.) THM 6-2415WI 24 VDC 250 mA 89 % THM 6-2421WI +5 VDC 600 mA 5 VDC 600 mA 85 % THM 6-2422WI +12 VDC 250 mA 12 VDC 250 mA 89 % THM 6-2423WI +15 VDC 200 mA 15 VDC 200 mA 89 % 3.3 VDC 1 800 mA 83 % THM 6-4810WI THM 6-4811WI 5 VDC 1 200 mA 87 % THM 6-4812WI 12 VDC 500 mA 88 % THM 6-4813WI 18 - 75 VDC 15 VDC 400 mA 89 % (48 VDC nom.) THM 6-4815WI 24 VDC 250 mA 88 % THM 6-4821WI +5 VDC 600 mA 5 VDC 600 mA 85 % THM 6-4822WI +12 VDC 250 mA 12 VDC 250 mA 88 % THM 6-4823WI +15 VDC 200 mA 15 VDC 200 mA 87 % Page 1 / 5 www.tracopower.com May 5, 2021THM 6WI Series, 6 Watt Options - Optional models with alternative pinning on demand - Optional models with adjustable output (backorder with MOQ - Optional models with remote-control function non stocking item) - Optional models with adjustable output and remote-control function Input Specifications Input Current - At no load 5 Vin models: 20 mA typ. 24 Vin models: 6 mA typ. 48 Vin models: 4 mA typ. Surge Voltage 5 Vin models: 16 VDC max. (3 s max.) 24 Vin models: 50 VDC max. (3 s max.) 48 Vin models: 100 VDC max. (3 s max.) Under Voltage Lockout 5 Vin models: 3 VDC min. / 4 VDC typ. / 4.4 VDC max. 24 Vin models: 7 VDC min. / 8 VDC typ. / 8.8 VDC max. 48 Vin models: 15 VDC min. / 16 VDC typ. / 17.5 VDC max. Recommended Input Fuse 5 Vin models: 2 500 mA (slow blow) 24 Vin models: 1 250 mA (slow blow) 48 Vin models: 630 mA (slow blow) (The need of an external fuse has to be assessed in the final application.) Input Filter Internal Pi-Type Output Specifications Output Voltage Adjustment 10% to +20% (15 & 24 Vout single models) 10% (other models) (Only for optional models with adjustable output) (By external trim resistor) See application note: www.tracopower.com/overview/thm6wi Output power must not exceed rated power Voltage Set Accuracy 1% max. Regulation - Input Variation (Vmin - Vmax) single output models: 0.2% max. dual output models: 0.5% max. - Load Variation (0 - 100%) single output models: 0.2% max. dual output models: 1% max. (Output 1) 1% max. (Output 2) - Cross Regulation dual output models: 5% max. (25% / 100% asym. load) Ripple and Noise - single output 3.3 Vout models: 30 mVp-p typ. (w/ 10 F X7R) (20 MHz Bandwidth) 5 Vout models: 30 mVp-p typ. (w/ 10 F X7R) 12 Vout models: 40 mVp-p typ. (w/ 10 F X7R) 15 Vout models: 40 mVp-p typ. (w/ 10 F X7R) 24 Vout models: 50 mVp-p typ. (w/ 4.7 F X7R) - dual output 5 / -5 Vout models: 30 / 30 mVp-p typ. (w/ 10 F X7R) 12 / -12 Vout models: 40 / 40 mVp-p typ. (w/ 10 F X7R) 15 / -15 Vout models: 40 / 40 mVp-p typ. (w/ 10 F X7R) Capacitive Load - single output 3.3 Vout models: 2 100 F max. 5 Vout models: 1 500 F max. 12 Vout models: 260 F max. 15 Vout models: 210 F max. 24 Vout models: 75 F max. - dual output 5 / -5 Vout models: 860 / 860 F max. 12 / -12 Vout models: 150 / 150 F max. 15 / -15 Vout models: 110 / 110 F max. Minimum Load Not required Temperature Coefficient 0.02 %/K max. All specifications valid at nominal voltage, full load and +25C after warm-up time unless otherwise stated. Page 2 / 5 www.tracopower.com May 5, 2021