MWA080Series 80WattMedicalDesktopPowerSupply HighEfficiency HighPowerDensity4.8W/in3 LifetimeExpectation>5years HoldupTime>24msatfullload EISA,CECCompliant FloatingOutput MedicalApprovalEN606011ClassI3rdEdition ElpacPart Output Output Peak Total Typical 1 2 3 Number Voltage Current Current Regulation Efficiency MWA080012A12A 12.0V 6.7A 8.0A 5% 88% MWA080015A12A 15.0V 5.3A 6.4A 5% 89% MWA080018A12A 18.0V 4.4A 5.3A 5% 89% MWA080024A12A 24.0V 3.3A 4.0A 5% 89% Notes 1Maximumpeakload(120W)lasting500mswithamaximum10%dutycycle. 2Includesinitialsetting,lineregulation,loadregulation,andthermaldrift. 3Typicalat115VAC(includingoutputcable).Input InputVoltage 85264VAC100240VACNominal InputFrequency 4763Hz InputCurrent <2.3Arms InrushCurrent <37Aat230VACcoldstart ZeroLoadPowerConsumption <0.5W EarthLeakageCurrent(Typical) <150A 132VAC 60H <250A 264VAC 60H PatientLeakageCurrent <50A 132VAC 60H <80A 264VAC 60H Output OutputVoltage SeeTable TotalRegulation +/5% MinimumLoad Nominimumloadrequired StartUpDelay <1s HoldUpTime >24msatanyinputvoltage Ripple&Noise <1%pkpk* OverVoltageProtection 110135% OverTemperatureProtection ActiveRecoverableplusPassiveNonRecoverable OverCurrentProtetion 120180% ShortCircuitProtection shutdown,autorestart(hiccupmode) Notes *Rippleandnoisemeasuredwith20MHzbandwidth10Ftantalumcapacitorinparallelwitha0.1Fceramiccapacitor. General Efficiency AvgEfficiency88.9% 115VAC90.6% 230VAC MTBF min.200,000hoursdemonstrated Size 5.19 (132mm)x3.07 (78mm)x1.34 (34mm) Weight 1.1lbs(0.50kg) PowerDensity 4.8W/in3